Test Code LAB116 PSA, Screening
Clinical Information
Do not predict disease recurrence solely on serial PSA values. Do not interpret levels of PSA as absolute evidence of the presence or the absence of malignant disease. Digital rectal examination generally does not increase normal prostate-specific antigen (PSA) values. However, cystoscopy, urethral instrumentation, and prostate biopsy may increase PSA levels. This assay is intended to be used as an aid in the detection of prostate cancer and as an aid in the management (monitoring) of prostate cancer patients, in accordance with current clinical practice guidelines. Specimens obtained from patients undergoing prostate manipulation, especially needle biopsy and transurethral resection, may show erroneously high results. Prostate cancer patients under treatment with anti-androgens and LHRH agonists may exhibit markedly reduced levels of PSA. Also, men treated for benign prostatic hyperplasia with inhibitors of 5α‑reductase (finasteride) may demonstrate a significant reduction in PSA levels compared to values prior to treatment.
For assays employing antibodies, the possibility exists for interference by human anti-animal antibodies (i.e., heterophile antibodies) in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, (eg, HAMA), that interfere with immunoassays. This may falsely elevate or falsely decrease the results.
Synonym
- PSA
Specimen Required
Preferred Container/Tube: Gold SST Top
Acceptable Container/Tube: Gold SST Top
Specimen Volume: 3 mL
Specimen Minimum Volume: 1 mL
Specimen Stability Information
Ambient: 8 hours
Refrigerated: 5 days
Rejected Due To
- Gross hemolysis
Reference Values
≤4.00 ng/mL
Method Description
Two-site sandwich immunoassay chemiluminescent.
Performing Lab
Clinical Lab UH
Clinical Lab University East
Clinical Lab James
Clinical Lab West Campus
Day(s) Performed
Monday through Sunday
Report Available
Same day/1 day
Reporting Name
PSA Screen
CPT Code Information
84153
G0103 (if appropriate)
LOINC Code Information
2857-1