Test Code LAB523 Estradiol, Enhanced
Clinical Information
The drug Fulvestrant (FASLODEX®) may cause falsely elevated estradiol results. For patients being treated with Fulvestrant, an alternate method that is not expected to show cross-reactivity to Fulvestrant, such as liquid chromatography-mass spectrometry (LC-MS), should be used. For assays employing antibodies, the possibility exists for interference by human anti-animal antibodies (i.e., heterophile antibodies) in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, (eg, HAMA), that interfere with immunoassays. This may falsely elevate or falsely decrease the results.
Synonym
- EE2B
Specimen Required
Preferred Container/Tube: Gold Top
Acceptable Container/Tube: Gold Top
Specimen Volume: 3 mL
Specimen Minimum Volume: 1 mL
Specimen Stability Information
Ambient: 20 hours
Refrigerated: 5 days
Rejected Due To
- Gross hemolysis
Interpretation
Male: 19+ years: <11.8-39.8
Menstruating Females:
- Follicular phase: 19.5-144.2
- Midcycle peak: 63.9-356.7
- Luteal phase: 55.8-214.2
Post-menopausal: <11.8-32.2
This test is not recommended for patients receiving Fulvestrant (Falsodex) due to possible false elevations. EE2 levels vary widely throughout the menstrual cycle.
Method Description
Competitive assay format. The endogenous estradiol contained in a sample is released from its binding proteins by a releasing agent. Then, a sheep monoclonal anti-estradiol antibody labeled with acridinium ester is added to bind available estradiol. Finally, an estradiol derivative capture solid phase is added to the reaction to compete with estradiol for the binding of the acridinium-labeled antibody. After washing, acid and base are dispensed to initiate the chemiluminescent reaction.
Performing Lab
Clinical Lab UH
Day(s) Performed
Monday through Sunday
Report Available
Same day/1 day
Reporting Name
Estradiol, Enhanced
CPT Code Information
82670
LOINC Code Information
2243-4