Test Code BG2 Venous Blood Gas
Synonym
- GASV5
Profile Information
| Reporting Name | Available Separately | Always Reported |
|---|---|---|
| pH, Venous | No | Yes |
| pCO2, Venous | No | Yes |
| pO2, Venous | No | Yes |
| HCO3, Venous | No | Yes |
| O2 Saturation, Venous | No | Yes |
| Base Excess | No | Yes |
Specimen Required
Preferred Container/Tube: Dry Lithium Heparin Syringe
Acceptable Container/Tube: Dry Lithium Heparin Syringe
Specimen Volume: Full syringe
Specimen Minimum Volume: Full syringe
Specimen Stability Information
Ambient: 30 minutes
Rejected Due To
- Unacceptable container type
- Quantity not sufficient
- Clotted specimens
- Frozen specimens
- Received >30 minutes from collection time
Reference Values
pH, Venous
7.32-7.43
pCO2, Venous
36-52 mm Hg
PO2, Venous
Venous pO2 is not recommended for the evaluation of oxygen status, clinical correlation is recommended.
HCO3, Venous
22-29 mmol/L
sO2, Venous
Venous sO₂ lacks a defined reference range due to wide physiologic variability; please interpret clinically.
Base Excess
-3.0-3.0 mmol/L
Cautions
pH: Contamination of sample with ambient air will falsely increase the pH.
Method Description
pO2: Amperometry
pH, pCO2: Potentiometry
Base excess: HCO3- - 24.8 + 16.2 x (pH - 7.4)
HCO3: Log (HCO₃ (c)) = pH + log(pCO₂) – 7.608 mmol/L
sO2: Oximetry
Performing Lab
Clinical Lab UH
Clinical Lab West Campus
Respiratory Therapy East
Respiratory Therapy UH
Day(s) Performed
Monday through Sunday
Reporting Name
Venous Blood Gas
CPT Code Information
82805
LOINC Code Information
11556-8
11558-4
11557-6
11556-8
1959-6
2713-6
11555-0