Test Code IMML18 CA 15-3N

Clinical Information

This test is not useful as a cancer screening test. Serum markers are not specific for malignancy and values may vary by method. Values obtained with different assay methods cannot be used interchangeably. For assays employing antibodies, the possibility exists for interference by human anti-animal antibodies (i.e., heterophile antibodies) in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, (eg, HAMA), that interfere with immunoassays. This may falsely elevate or falsely decrease the results.

Synonym

  • CA153

Specimen Required

Preferred Container/Tube: Mint Green Top
Acceptable Container/Tube: Mint Green Top, Gold SST Top
Specimen Volume: 3 mL
Specimen Minimum Volume: 1 mL

Specimen Stability Information

Ambient: 8 hours
Refrigerated: 5 days

Rejection Due To

  • Gross hemolysis
  • Quantity not sufficient
  • Quality not acceptable

Reference Values

FEMALE MALE
0.0-32.4 U/mL 0.0-32.4 U/mL
(use not defined)

 

Method Description

Two-site sandwich immunoassay chemiluminescent.

Performing Lab

Clinical Lab UH

Day(s) Performed

Monday through Sunday

Report Available

Same day/1 day

Reporting Name

CA 15-3

CPT Code Information

86300

LOINC Code Information

6875-9