Test Code LAB527 Insulin

Clinical Information

Insulin autoantibodies in human serum may interfere and cause discordant results. This assay has 100% cross-reactivity with recombinant human insulin (Novolin R and Novolin N), analogues of injectable insulin (i.e., insulin lispro, insulin detemir, insulin aspart, and insulin glargine). For assays employing antibodies, the possibility exists for interference by human anti-animal antibodies (i.e., heterophile antibodies) in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, (e.g., HAMA), that interfere with immunoassays. This may falsely elevate or falsely decrease the results.

Synonym

  • INSL

Specimen Required

Preferred Container/Tube: Gold SST Top
Acceptable Container/Tube: Gold SST Top
Specimen Volume: 3 mL
Specimen Minimum Volume: 1 mL

Specimen Stability Information

Refrigerated: 24 hours

Rejected Due To

  • Gross hemolysis
  • Patient not fasting

Reference Values

3.0-25.0 uIU/mL

Method Description

Two-site sandwich chemiluminescent immunoassay 

Performing Lab

Clinical Lab UH

Day(s) Performed

Monday through Friday (excludes OSU holidays)

Report Available

2 to 4 days

Reporting Name

Insulin

CPT Code Information

83525

LOINC Code Information

20448-7