Test Code LAB527 Insulin
Clinical Information
Insulin autoantibodies in human serum may interfere and cause discordant results. This assay has 100% cross-reactivity with recombinant human insulin (Novolin R and Novolin N), analogues of injectable insulin (i.e., insulin lispro, insulin detemir, insulin aspart, and insulin glargine). For assays employing antibodies, the possibility exists for interference by human anti-animal antibodies (i.e., heterophile antibodies) in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, (e.g., HAMA), that interfere with immunoassays. This may falsely elevate or falsely decrease the results.
Synonym
- INSL
Specimen Required
Preferred Container/Tube: Gold SST Top
Acceptable Container/Tube: Gold SST Top
Specimen Volume: 3 mL
Specimen Minimum Volume: 1 mL
Specimen Stability Information
Refrigerated: 24 hours
Rejected Due To
- Gross hemolysis
- Patient not fasting
Reference Values
3.0-25.0 uIU/mL
Method Description
Two-site sandwich chemiluminescent immunoassay
Performing Lab
Clinical Lab UH
Day(s) Performed
Monday through Friday (excludes OSU holidays)
Report Available
2 to 4 days
Reporting Name
Insulin
CPT Code Information
83525
LOINC Code Information
20448-7