Sign in →

Test Code YQFTB M Tuberculosis By Quantiferon

Clinical Information

QuantiFERON-TB Gold Plus is an interferon-gamma release assay that detects a cell-mediated immune response to Mycobacterium tuberculosis antigens in whole blood. The antigens used in the assay are absent from BCG vaccine strains and most nontuberculous mycobacteria; therefore, prior BCG vaccination is not expected to cause a positive result. Cross-reactivity may occur with Mycobacterium kansasii, Mycobacterium marinum, or Mycobacterium szulgai.
This test is an indirect aid in identifying M. tuberculosis infection and should be interpreted with the patient’s exposure risk, clinical findings, imaging, and other diagnostic results. It cannot distinguish latent tuberculosis infection from active tuberculosis disease and should not be used alone to diagnose or exclude active tuberculosis. The predictive value of a negative result in immunosuppressed patients has not been established.

Synonym

  • QFTB
  • QFT
  • M. Tuberculosis Antigen

Specimen Required

Preferred Container/Tube: Quantiferon Kit
Acceptable Container/Tube: Quantiferon Kit
Specimen Volume: 1 mL
Specimen Minimum Volume: 1 mL

Specimen Stability Information

Ambient: 16 hours
Refrigerated: 28 days

Specimen Collection Information

DO NOT FILL TUBES WITH A SYRINGE. Tubes should be at 17-25C (63-77F) at the time of blood filling. Collect 1 ml of blood into each of the 4 tubes.

 

Order of draw:

  1. Gray (Nil)
  2. Green (TB1 )
  3. Yellow (TB2)
  4. Purple (Mitogen)

Tubes fill slowly. Keep each tube on the needle for 2-3 seconds after tube appears to have completed filling. When tube is upright, blood level must be within the small black mark on the label on all 4 tubes. Over filling or under filling will cause the test to be rejected. Immediately shake the tubes 10 times, just firmly enough to ensure inner surface of tube is coated in blood (to solubilize antigens on tube walls). Do not vigorously shake them. Over-energetic shaking could lead to aberrant results. Ensure the color name printed on the label is placed on corresponding colored tube, i.e., purple label to purple tube, etc.  Make sure the QuantiFERON band and entire specimen volume is visible. WRITE THE COLLECT TIME ON ALL THE TUBES. Maintain the tubes at room temperature.

Rejection Due To

  • Improperly labeled specimens
  • Unacceptable container type
  • Microbially contaminated specimens
  • Non-venipuncture collections
  • Samples refrigerated before incubation and centrifugation

Interpretation

Negative

Method Description

Chemiluminescent immunoassay

Performing Lab

Clinical Lab UH

Day(s) Performed

Monday through Friday (excludes OSU holidays)

Report Available

4 to 6 days

Reporting Name

M Tuberculosis By Quantiferon

CPT Code Information

86480

LOINC Code Information

45323-3
46216-8 x4